To ensure the validity of your pulmonary testing results in the field, measurements from your spirometer should be regularly checked against a precise calibration syringe. The ATS/ERS Task Force standards, published Spring 2005, require daily calibration checks of flow and volume spirometers using a 3-L calibration syringe with an accuracy of ±15 mL or ±0.5% of the full scale.
Futuremed’s 3.0 liter calibration syringe is a reliable precision tool for quickly and easily verifying and fine-tuning your Futuremed spirometer and all other pulmonary function testing equipment. It is built with an accuracy of 0.5% of the full scale to meet the ATS/ERS recommendations.
Order number 39250 – Certified 3 Liter Calibration Syringe